Configuration of the RS232 output of INDI-5250 or INDI-00

  1. Go to menu SETUP:

    Press  , Enter Fn49 via keys   and 
    Validate via   Display SETUP
    Validate via 

  2. Modification of communication parameters
    Press (1x)  on  to get SETUP2, pthen press 
    Verify all the parameters hereunder by pressing  

    2.t = 02
    2.c = 65
    2.l = 01
    2.r = 00
    2.a = 00
    2.f = 00
    2.e = 1
    2.1 = 0
    2.2 = 0
    2.3 = 0
    2.4 = 0
    2.5 = 0
    2.6 = 0
    2.7 = 0
    2.8 = 0
    2.d = 17
    SETUP2
    Press  to quit

  3. Modification of the display update rate:

    Press  to display « Par »
    Press  several times until « 6.P » appears and encode « 14 » by using 
    Press  to validdate
    Press  to quit.
                                                                                                                               
  4. Save changes:

    Press 2x  to get « STORE » and validate by 

  5. Cable to usr:

    RS232 cable used between the output of the INDI-5250’ and a PC must be crossed (TX and RX).



  6. Communication parameters:                                                                                                                                                                                                           

    2400 bauds – ATTENTION

    7 data bits + 1 parity bit

    or

    8 data bits + 0 parity bit

    1 stop bit