STANDARD REFERENCE FORCE TRANSDUCERS IN TENSION AND COMPRESSION
2715-ISO
Standard reference force transducers specially designed according to the ISO 376 standard (classes 1, 0,5 and 00).
o Wide range of capacities
o Tension and / or compression (universal)
o Compact design
o Complete range of load accessories
o Protection class: IP65
o Material: alloy steel (stainless steel available on request)
o Cable length: see drawing (other lengths available on request)
1 | 05 | 00 | |
---|---|---|---|
Relative reversibility error (% M.V.**) | <± 0.3 | <± 0.15 | <± 0.07 |
Relative repeatability error with rotation (% M.V.**) | <± 0.2 | <± 0.1 | <± 0.05 |
Relative repeatability error without rotation (% M.V.**) | <± 0.1 | <± 0.05 | <± 0.025 |
Stabilization time after power excitation supply (s) | 200...600 | 200...600 | 200...600 |
Creep error over 30 min. (% F.S.*)
|
<± 0.1 | <± 0.05 | <± 0.025 |
Reference temperature (°C) | 20 | 20 | 20 |
Compensated temperature range (°C) | -10...+45 | -10...+45 | -10...+45 |
Service temperature range (°C)
|
-25...+70 | -25...+70 | -25...+70 |
Storage temperature range (°C)
|
-50...+85 | -50...+85 | -50...+85 |
Temperature coefficient of the sensitivity (% F.S.*/10°C) | <± 0.035 | <± 0.035 | <± 0.015 |
Temperature coefficient of zero signal (% F.S.*/10°C) | <± 0.03 | <± 0.03 | <± 0.023 |
Zero balance (mV/V)
|
± 0.02 | ± 0.02 | ± 0.02 |
Nominal sensitivity (mV/V)
|
2 | 2 | 2 |
Input resistance (ohm(s)) | 350 ± 2 | 350 ± 2 | 350 ± 2 |
Output resistance (ohm(s)) | 350 ± 2 | 350 ± 2 | 350 ± 2 |
Insulation resistance (50 V) (Mohm(s)) | > 5000 | > 5000 | > 5000 |
Reference excitation voltage (VDC)
|
10 | 10 | 10 |
Permissible nominal range of excitation voltage (VDC) | 3...12 | 3...12 | 3...12 |
Safe load limit (% F.S.*) | 110 | 110 | 110 |
Breaking load (% F.S.*)
|
>300 | >300 | >300 |
* F.S. : Full Scale.
** : except thermal drifts.
Specifications subject to change without notice.